Completely remove the prosthesis of the implant that needs to be removed.
Select an appropriate size of Perfect Guide.
By using the PG Hand Driver, tighten the Perfect Guide To the fixture in a clockwise direction until it touches the screw inside the fixture.
Select an appropriate size of Reverse Drill and connect it to the contra angle of the surgical engine.
Insert the Reverse Drill into the Perfect Guide attached on the fixture.
By drilling counter clockwise at a speed of approximately 2,000rpm, drill 1~2mm into the surface of the fractured screw.
Check the marking line(in 1mm intervals) of the drill body while drilling with adequate irrigation.
Remove the Perfect guide connected to the fixtureSelect a Screw Remover that fits to the hole created by the Reverse Drill.Connect the Screw Remover to the contra angled of surgical engine and drill slowly(less than 80rpm) while pressing down with enough force to remove the fractured screw.