Minimize bone loss while removing implant

Surgical drills and fixture drivers for fixture implantation are included.

Improvement of fixation and increase clinician’s comfort when placing implants by widening cutting edge area 2 times compared to IS-II Active.

No need to use a trephine drill

Use Fixture remover screw, Fixture remover and Torque wrench

Prevents loss of cortical bone thread line

Surgical drills and fixture drivers for fixture implantation are included.

Improvement of fixation and increase clinician’s comfort when placing implants by widening cutting edge area 2 times compared to IS-II Active.

FR Kit only rotates failed implant to remove

Fixture with identical diameter as removed one can be implanted with remaining bone thread line

Applicable to any type of implants with a screw hole

* Option Tool

Improvement of fixation and increase clinician’s comfort when placing implants by widening cutting edge area 2 times compared to IS-II Active.

FR Kit is equipped with fixture remover screw of various sizes.

Available from M1.4 to M2.5 of screw hole size

1.Fixture Remover Screw

Length(mm) / Product name
Screw size
Recommended Torque(Ncm)
Product name
M1.4
60
FRS14
M1.6
80
FRS16
M1.8
80
FRS18
1-72UNF
80
FRS172
M2.0
80
FRS20
M2.53
80
FRS25
*”M” means a metric screw, and the number stands for the  
 size of the external diameter of the screw

2.FR Screw Hex Driver

Length(mm) / Product name
Length
Product name
7mm
HDF1607
12mm
HDF1612
17mm
HDF1617

3.Fixture Remover

Length(mm) / Product name
Type
Product name
Product name
Narrow
15mm
FRS315
Narrow
20mm
FRS320
Regular
15mm
FRS415
Regular
20mm
FRS420
Wide(Ø5)
15mm
FRS515
Wide(Ø5)
20mm
FRS520
Wide(Ø6~8)
15mm
FRS615
Wide(Ø6~8)
20mm
FRS620

4.Torque Control Device

Length(mm) / Product name
Type
Product name
Body
TW80400
Slider
FRCHT

5.Fixture Remover

Length(mm) / Product name
Type
Length
Type
Product name
Narrow
FRS315 or FR320
FRS14
200
Narrow
FRS315 or FR320
FRS16
250
Narrow
FRS315 or FR320
FRS18/FRS172
200
Narrow
FRS315 or FR320
FRS20
300
Regular
FR315 or FR320
FRS18/FRS172
350
Regular
FR315 or FR320
FRS20
300
Regular
FR315 or FR320
FRS18/FRS172
350
Regular
FR315 or FR320
FRS20
300
Wide(Ø5)
FR415 or FR420
FRS20
350
Wide(Ø5)
FR415 or FR420
FRS20
400
Wide(Ø5)
FR415 or FR420
FRS20
350
Wide(Ø5)
FR415 or FR420
FRS25
400
Wide(Ø6~8)
FR515 or FR520
FRS20
350
Wide(Ø6~8)
FR515 or FR520
FRS25
400
Wide(Ø6~8)
FR515 or FR520
FRS20
350
Wide(Ø6~8)
FR515 or FR520
FRS25
400

Crestal approach

1
Failed Implant
  • Completely remove the prosthesis of the implant that needs to be removed.

2
Fixture Remover Screw
  • Select an appropriate size of Fixture Remover Screw.

  • Connect the Fixture Remover Screw to the FR Screw Hex Driver and tighten it clockwise.

3
Fixture Remover
  • Select an appropriate size of Fixture Remover.

  • Connect the selected Fixture Remover and tighten it counterclockwise until it is completely tightened to the Fixture Remover Screw that is fixed in the implant.

  • Using a Torque Wrench, further apply torque to the Fixture Remover(counter clockwise) until the implant is finally unscrewed from the site.

4
Clear